Meridian-sphere OPM simulation

This is a separate, stationary-object forward simulation. Production orthogonal deskew and PSF generation are unchanged. The forward check below does not use deconvolution; the subsequent GC comparison uses the saved noisy camera data.

Object and optics

The ground truth is a hollow sphere with 10 um meridian-centerline diameter and 1 um diameter fluorescent tubes. Its outer diameter is therefore 11 um. Twelve equally spaced pole-to-pole meridians form six full great circles through the Z axis. Tubes join at the poles; overlapping fluorescence is a union, so the poles are not artificially brighter because multiple tubes overlap. The interior and gaps between meridians contain no fluorophores.

The object is defined first on an isotropic 50 nm Cartesian grid, finer than the 115 nm camera pitch and its 57.5 nm projected Z spacing at 30 degrees. The optical PSF is the vectorial 100x silicone model, with configurable NA and wavelength (defaults NA 1.35, 637 nm; immersion RI 1.4, sample RI 1.38). Optical support is +/-2.4 um in Z and +/-1.2 um in XY. Both simulation routes use this same finite-support PSF. There is no empirical taper.

Independent forward routes

OPM route: interpolate the fine Cartesian object and PSF into instrument coordinates. The fine instrument grid has 0.1 um scan spacing and 57.5 nm camera-row/column spacing. Convolve the object with the normalized skewed PSF on that fine grid. Only then select the requested acquired scan planes and sample/average camera pixels. The transformations are

X = camera_column * pixel_pitch
Y = scan_index * scan_step + camera_row * pixel_pitch * cos(angle)
Z = camera_row * pixel_pitch * sin(angle)

All grids have explicit centered coordinates in microns. Optical halos and the complete object are retained before convolution, which uses zero exterior fluorescence rather than FFT wraparound.

Cartesian microscope route: independently convolve the ground truth and Cartesian optical PSF on the fine Cartesian grid, then digitally sample it. For the comparison, evaluate it at the physical coordinates of the deskewed output. No registration, fitted gain, or alignment optimization is applied.

There is also a direct check before deskew: sample the Cartesian convolution on the raw OPM grid and compare it with the independently computed skewed convolution. This distinguishes forward-model sampling error from later deskew interpolation differences.

Production deskew applies a constant-field gain of 2*pixel/scan_step. Since the simulator's clean arrays represent fluorescence density, multiply its output by the reciprocal of that known factor for the comparison. This does not change production deskew or fit image brightness. Deskew Z downsampling is disabled for these comparisons; output voxels are 115 nm isotropic.

Camera sampling and noise

Start with --camera-samples 1, which samples camera-pixel centers without noise. After geometry validation, --camera-samples 3 averages a 3x3 set of uniform subpixel centers within each detector pixel. This approximates binning the finer irradiance field over the camera pixel area. The OPM detector plane is tilted; the Cartesian microscope detector averages in laboratory XY. Their pixel-area responses need not be identical.

Convert the averaged intensity to expected detected photoelectrons using --peak-electrons, add --background-electrons, draw Poisson counts, and then add Gaussian read noise with --read-noise in electrons RMS. Noise is applied once per camera pixel after averaging, not once per interpolation subpixel. Both microscopes use the same exposure gain and independent noise draws. The gain is based on the peak of the fine Cartesian optical image, not fitted to the sampled OPM image. Background-subtracted density is clipped at zero before deskew, introducing a small positive background bias; original electron-valued raw data are also saved. ADC quantization, saturation, motion, and rolling shutter timing are not included.

Reproduce

Run from the repository root:

uv run python -m scripts.opm_simulation --output diagnostics/meridian_sphere --scan-steps 0.2 0.8

Then add camera integration and noise:

uv run python -m scripts.opm_simulation --output diagnostics/meridian_sphere_camera --scan-steps 0.2 0.8 --camera-samples 3 --peak-electrons 500 --background-electrons 2 --read-noise 1.5

Geometry and optics can be changed with --diameter, --tube-diameter, --meridians, --angle, --pixel-size, --na, --wavelength, --fine-spacing, and --fine-scan-step. Wavelength and all spatial inputs are in microns. --seed controls camera noise. Finer grids cost more memory/time.

Each run exports fine ground truth and both PSFs. Each acquisition directory contains raw skewed data, deskewed data, the Cartesian microscope reference, the noise-free reference, and sampled truth. Cartesian comparison volumes are OME-TIFFs with physical voxel sizes, as are the fine Cartesian truth and PSF. Raw TIFF Z denotes scan index, not Cartesian Z. simulation.json records the actual axes, physical origins, parameters, and quantitative comparisons. comparison.png shows maximum projections, sections.png shows the central planes through the hollow sphere, and raw_instrument.png shows the oblique data before deskewing.

Render the saved volumes without rerunning optics:

uv run --with matplotlib python -m scripts.plot_opm_simulation diagnostics/meridian_sphere/scan_0.2um

Results and checks

For the noiseless default object, the independent forward routes agree to 1.96% relative L2 error and 0.999913 correlation. After unchanged orthogonal deskew:

Acquired scan step Error vs Cartesian microscope Correlation Total signal ratio
0.2 um 2.81% 0.99944 1.00037
0.8 um 8.84% 0.99222 1.00028

With 3x3 pixel integration, 500 peak electrons, 2 background electrons and 1.5-electron RMS read noise, errors against the noise-free Cartesian reference are 4.88% and 9.66%, respectively. Comparisons against an independently noisy Cartesian image have larger errors (8.47% and 11.87%).

Errors/correlations use voxels above 1% of the reference peak, excluding the large empty background; total-signal ratios use the complete compared volume. Projection figures use shared intensity scales and include the projection of the absolute volumetric error, rather than differences between independent maximum projections.

The original 23 numerical cases covered exact tube geometry and empty interior, pixel-area quadrature, Poisson/read-noise moments, the two convolution routes, physical sphere centering, deskew/reference agreement at two scan steps, and fine-grid PSF sampling-convergence checks. Current in-memory simulation checks are unit tests. Integration tests must exercise simulated data from disk input to verified disk output, as described in the testing guide. These checks validate the specified stationary forward model; they do not explain motion or residual tilt in a live acquisition.

GC after the complete camera simulation

The required order is fine object -> optical convolution -> oblique camera sampling and pixel integration -> photon/read noise -> GC -> orthogonal deskew. Noiseless geometry-check data are rejected by the GC comparison. Earlier exploratory GC runs on those noiseless data were withdrawn and are not evidence for changing production code.

The comparison reads the existing raw_electrons.tif, subtracts the recorded background, clips negative calibrated counts as in the simulation's camera calibration, and runs the existing GC solver. It does not regenerate the image or draw another noise realization. An SHA-256 of the input electron-data file is saved with the results. Conversion back to density uses the original camera gain, not a fitted normalization. For older exports without that gain recorded, its exact definition is recalculated from the saved fine truth and PSF and checked against the saved calibrated camera image. New exports record it.

The deconvolution PSF is sampled from the same fine Cartesian optical PSF at the reconstruction spacing and includes camera-pixel integration. Native GC uses the acquired scan step. Optional missing-plane GC uses the finer requested step, preserves the original endpoints, and passes that step to unchanged orthogonal deskew. The Cartesian microscope reference is also deconvolved with its corresponding pixel-integrated PSF.

uv run python -m scripts.deconvolve_opm_simulation diagnostics/meridian_sphere_camera/scan_0.2um
uv run python -m scripts.deconvolve_opm_simulation diagnostics/meridian_sphere_camera/scan_0.8um
uv run python -m scripts.deconvolve_opm_simulation diagnostics/meridian_sphere_camera/scan_0.8um --scan-upsample 2

Outputs are in decon_native or decon_upsample2 subdirectories and can be rendered with the same plotting module. Unlike the optical-image comparisons above, the following relative L2 errors include all voxels and compare with the unblurred ground truth, including false fluorescence outside the tubes:

Acquisition / reconstruction Before deconvolution After deconvolution
OPM 0.2 / 0.2 um 72.20% 38.68%
OPM 0.8 / 0.8 um 73.44% 42.70%
OPM 0.8 / 0.4 um 73.44% 39.31%
OPM 1.2 / 0.4 um 74.78% 38.54%
Cartesian microscope 72.60% 39.35%

The two 0.8 um runs use exactly the same input electron data. These are noisy, finite-resolution reconstructions, not exact recovery of the binary tube boundaries. No production deconvolution, PSF-generation or deskew changes were needed for these comparisons.

The 1.2 um acquisition uses the same optics, 3x3 camera integration, 500 peak electrons, 2 background electrons, 1.5-electron read noise and seed 42. Its 37 measured planes reconstruct to 109 planes at 0.4 um using factor 3. Reproduce it:

uv run python -m scripts.opm_simulation --output diagnostics/meridian_sphere_camera_1p2 --scan-steps 1.2 --camera-samples 3 --peak-electrons 500 --background-electrons 2 --read-noise 1.5 --seed 42
uv run python -m scripts.deconvolve_opm_simulation diagnostics/meridian_sphere_camera_1p2/scan_1.2um --scan-upsample 3

Against the deconvolved Cartesian reference, this reconstruction has 14.15% relative error and 0.98699 correlation on the reference mask, compared with 11.07% and 0.99126 for the 0.8 / 0.4 um reconstruction. Its slightly lower error against unblurred truth in this single noise realization does not establish an advantage for coarser acquisition.